Advanced life support solutions

HeartStart MRx

Monitor/Defibrillator for emergency care

Εύρεση παρόμοιων προϊόντων

Rugged, reliable, and easy-to-use, the HeartStart MRx with the DXL 12-Lead ECG algorithm, provides industry-leading patient monitoring capabilities, exceptional diagnostic measurements, proven resuscitation therapies, and CPR guidance.

Specifications

Physical Dimensions
Physical Dimensions
Size with external paddles
  • 12.4" (W) x 7.7" (D) x 13.4" (H) (313 mm x 195 mm x 340 mm) "/mm
Weight
  • 13.2 lbs. (6 kg): base unit with 1 battery/pads and pads cable.
  • 13 lbs. (5.9 kg) with optional 75 mm strip chart printer.
  • Paddle tray and external standard paddles add less than 2.5 lbs. (1.1 kg).
  • Carrying case adds 4.1 lbs. (1.86 kg). lbs/kg
Defibrillation
Defibrillation
Waveform
  • Truncated Exponential Biphasic.Waveform parameters adjusted as a function of patient impedance.
Output Energy
  • Manual (selected): 1-10/15/20/30/50/70/100/120/150/170/200 Joules into a 50 Ohm load
  • AED Mode (single energy output): 150 Joules into a 50 ohm load. Joules/Ohm
Charge Time
  • Less than 5 seconds to 200 Joules with a new/ fully charged lithium ion battery at 25º C
Shock Delivery
  • Via multifunction defib electrode pads or paddles
Shock-to-Shock Cycle Time
  • Typically less than 20 seconds seconds
Minimum Patient Impedance Range
  • 15 Ohm (internal defibrillation); 25 Ohm (external defibrillation) Ohm
Maximum Patient Impedance Range
  • 180 Ohm Ohm
AED Mode
  • Shock advisory sensitivity and specificity meet AAMI DF-39 guidelines
Physical dimensions Defibrillator model HeartStart MRx (M3536A)
Physical dimensions Defibrillator model HeartStart MRx (M3536A)
Size without external paddles
  • 12.4" (W) x 7.7" (D) x 11.7" (H) (313 mm x 195 mm x 295 mm) "/mm
CPR feedback and measurement
CPR feedback and measurement
Compression Depth
  • Target: -1.50 to -2.00" (-38 to -51 mm) "/mm
Compression Rate
  • Target: 90 to 120 cpm cpm
Ventilation Volume
  • Graphic indicator: empty/one-third full/two-thirds full/full
Ventilation Rate(a)
  • Target for breaths delivered before 60 seconds since last compression: 6 to 16 vpm seconds/vpm
Ventilation Rate(b)
  • Target for breaths delivered beyond 60 seconds since last compression: 9 to 16 vpm seconds/vpm
SpO2 pulse oximetry
SpO2 pulse oximetry
Range
  • 0 to 100% %
Resolution
  • 0,01%
Alarm Range(a)
  • Low Limit: 50 to 99% (Adult/Pediatric) %
Alarm Range(b)
  • High Limit: 51 to 100% (Adult/Pediatric) %
Alarm Delay
  • 10 seconds
Noninvasive Blood Pressure
Noninvasive Blood Pressure
Waveform
  • Monophasic Truncated Exponential
Pressure Range(a)
  • Systolic: 40 to 260 mmHg mmHg
Pressure Range(b)
  • Diastolic: 20 to 200 mmHg mmHg
Current Pulse Amplitude
  • 10 mA to 160 mA (5 mA resolution); accuracy 10 mA to 50 mA ± 5 mA, 50 mA - 160 m A ± 10% mA
Pulse Width
  • 40 ms with ± 10% accuracy ms/%
Initial Pressure
  • Adult: 160 mmHg and Pediatric: 120 mmHg mmHg
Rate
  • 30 ppm to 180 ppm (10 ppm increments); accuracy ± 1.5% ppm
Maximum Pressure
  • 280 mmHg mmHg
Modes
  • Demand or Fixed Rate
Alarm Range(d)
  • Systolic high limit: 30 - 270 (Adult) and 35 - 180 (Pediatric)
Refractory Period
  • 340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm) msec/ppm
Alarm Range(e)
  • Systolic low limit: 30 - 265 (Adult) and 30 - 175 (Pediatric)
Alarm Range(f)
  • Diastolic high limit: 18 - 240 (Adult) and 18 - 150 (Pediatric)
Alarm Range(g)
  • Diastolic low limit: 10 - 240 (Adult) and 10 - 145 (Pediatric)
ECG and arrhythmia monitoring
ECG and arrhythmia monitoring
Input
  • Up to 4 ECG waves displayed and up to 3 ECG waves print simultaneously.
Leads
  • Lead I, II, or III obtained through 3-lead ECG cable and separate monitoring electrodes.
  • With 5-lead cable, obtain leads I/II/III/aVR/aVL/ aVF or V.
  • Pads ECG obtained through 2 multifunction defibrillation electrode pads.
Lead Fault
  • ‘Lead Off ’ message and dashed line displayed if an electrode or lead wire becomes disconnected.
Pads Fault
  • Dashed line displayed if a pad becomes disconnected.
Heart Rate Display
  • Digital readout on display 15 to 300 bpm with accuracy ±10% bpm/%
Heart Rate/Arrhythmia Alarms
  • HR/Asystole/VFIB/VTACH/VTACH/extreme tachycardia/extreme bradycardia/PVC rate.
ECG Size
  • 2.5/5/10/20/40 mm/mV autogain mm/mV
Environmental & physical requirements
Environmental & physical requirements
Solids/Water resistance
  • IP24
Operating temperature
  • 0º - 45º C (32º - 113º F) ° C/° F
Storage temperature
  • -20 - 70º C (-4º - 158º F) ° C/° F
Operating humidity
  • 0% to 95% relative %
Operating altitude
  • 0 to 15000 ft (0 to 4500 m) ft/m
Storage altitude
  • 0 to 15000 ft (0 to 4500 m) ft/m
Mechanical shock
  • Bump: IEC 68-2-29
  • Freefall: IEC 68-2-32
Operating vibration
  • MIL STD 810E 514.4 Category 6 Helicopter
General storage vibration
  • UH60 Non-Operating: IEC 68-2-6 Swept Sine Vibration and IEC 68-2-64 Random Vibration
Safety
  • Meets EN 60601-1, UL 2601-1/CSA C22.2 No. 601-1
Strip Chart Recorder
Strip Chart Recorder
Recorder(a)
  • Standard: 50 mm (paper width) thermal array printer mm
Recorder(b)
  • Optional: 75 mm (paper width) thermal array printer mm
Continuous ECG Strip
  • Prints primary ECG lead with event annotations and measurements in real-time or with 10-second delay
Auto Printing
  • Recorder can be configured to print marked events/charge/ shock and alarms
Reports
  • Event Summary, 12-Lead/Operational Check/ Configuration/Status Log and Device Information
Paper Size
  • 1.97” (50 mm) W by 100 ft. (30 m) L and 2.95” (75 mm) W by 100 ft. (30 m) L "/ft/mm/m
Battery
Battery
Type
  • 6.3 Ah, 14.8 V, rechargeable lithium ion Ah/V
Dimensions
  • 6.5" (H) x 3.8" (W) x 1.6" (D) (165 mm x 95 mm x 42 mm) "/mm
Weight
  • 1.6 lb. (0.73 kg) lb/kg
Charge Time
  • Approximately 3 hours from fully depleted to 100%. 90 minutes from fully depleted to 80% hours/%
Capacity(a)
  • At least 5 hours of continuous 12-lead ECG/SpO2 and CO2 monitoring.
  • With NBP every 15 minutes on one new fully charged battery. hours
Capacity(b)
  • At least 3.5 hours of continuous 12-lead ECG/SpO2 and CO2 monitoring.
  • With NBP every 15 minutes and pacing at 180 ppm at 160 mA on one new fully charged battery. hours
12-Lead ECG
12-Lead ECG
Input
  • 12-Lead cable: leads I/II/III/aVR/aVL/ aVF and V/C1-V/C6
Display View
  • All 12-lead ECG waves display simultaneously
Strip Recorder
  • All 12-leads print on the strip chart printer in 3x4 format
Transmission
  • CompactFlash data card; Bluetooth® dial-up Internet connection, Bluetooth® FTP
Display
Display
Dimensions
  • 8.4’’ diagonal (128 mm x 171 mm) "/mm
Type
  • TFT color LCD
Resolution
  • 480 x 640 pixels (VGA) VGA
Wave Viewing Time
  • 5 seconds (ECG) seconds
End-tidal CO2
End-tidal CO2
Range
  • 0 to 99 mmHg mmHg
Resolution
  • 1 mmHg (0.1 kPa) mmHg/kPa
Sample Size
  • 50 ml per minute ml
Alarm Range(h)
  • Low Limit: 10 to 95 mmHg (Adult/Pediatric) mmHg
Alarm Range(i)
  • High Limit: 20 to 100 mmHg (Adult/Pediatric) mmHg
Data storage
Data storage
Internal
  • 12 hours of continuous ECG waveforms and events. Plus 50 12-lead ECG reports
Data Card
  • 12 hours of continuous ECG waveforms and events.
  • Plus 50 12-lead ECG reports on a CompactFlash memory card